2-1
2. Theory of Operation
Principles of Operation
The NDIR (Non-Dispersive Infra-Red Spectroscopy)
measurement method is implemented in the DX6106
Optical Unit.
The device provides gases concentration measure-
ment based on the classical double channel optical
scheme (Fig. 2.1). One of the beams (measuring
channel) has the wavelength which is tuned to the
optical absorption line of the measured gas. The other
beam (reference channel) has the wavelength which is
out of the adsorption band of the measured gas.
Intensities of two light beams that passed through the
measured gas sampling cell are compared.
Theory of Operation
RMT LtdRMT Ltd
RMT LtdRMT Ltd
DX6106
Fig. 2.1. The principle of gas concentration
l
2
l
2
l
1
l
1
Light Emitters
Photoresistor
Gas Flow
Intensity
Wavelength
l
1
l
1
l
2
l
2
Intensity
Wavelength